B & K Precision 1672 Instruction Manual

Browse online or download Instruction Manual for Power supply units B & K Precision 1672. B & K Precision 1672 Instruction manual User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Triple Output Power Supply

Model 1672, 1673 Triple Output Power Supply INSTRUCTION MANUAL

Page 2

8 13) Master current indicator display (3-digit red 0.56” LED) 14) Series mode indicator LED 15) Parallel mode indicator LED 16) Slave voltage i

Page 3 - 1 Safety Summary

9 5 Operating Instructions 5.1 Setting the Output Voltage and Output Current 1. As per load requirement, calculate the voltage and maximum curr

Page 4

10 For example, if the load is such that the power supply is operating in the constant voltage mode, a regulated output voltage is provided. The out

Page 5 - Compliance Statements

11 5.3.3 Parallel Tracking Mode In this mode, the positive output terminal of the master gets connected internally to the positive output term

Page 6 - Table of Contents

12 5.3.4 Multiple Units in Series Mode For achieving higher voltage, two or more units can be connected in series (Max 240V only). The output

Page 7 - 3 Installation

13 5.3.5 Multiple Units in Parallel Mode Two or more units can be connected in parallel to obtain a higher current output (M

Page 8 - 3.3 Location

14 5.3.6 Fixed 5V/3A Output This is the standard 5V/ 3A power output provided for supplying the power to TTL logic circuits.

Page 9 - 4 Controls and Indicators

15 6 Specifications 1672 1673 Output Parameters Number of Outputs Three (two variable and one fixed) Range 0 to 32 VDC / 0 to 3 A (variable) and

Page 10 - 4.2 Rear Panel

16 General Power Requirements 115/230 V ± 10%, 50/60 Hz ± 10% Operating Temperature 50 °F to 104 °F (10 °C to 40 °C) Humidity 90% R.H. Temperature C

Page 11 - 5 Operating Instructions

17 • Do not obstruct the ventilation holes in the rear panel as this will increase the internal temperature. • Do not operate the instrument with

Page 13

18 8 Service Information Warranty Service: Please go to the support and service section on our website www.bkprecision.com to obtain an RMA #. Ret

Page 14

19 9 Limited One-Year Warranty B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will

Page 15

(Page intentionally left blank)

Page 16

Printed in Taiwan v112310

Page 17 - 6 Specifications

1 1 Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases of op

Page 18 - 7 Maintenance

2 WARNINGS AND CAUTIONS WARNING and CAUTION statements, such as the following examples, denote a hazard and appear throughout this manual. Follow a

Page 19 - 7.3 Fuse Replacement

3 Compliance Statements Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with se

Page 20 - 8 Service Information

4 Table of Contents 1 Safety Summary ...

Page 21 - 9 Limited One-Year Warranty

5 2 Introduction Description B&K Precision’s 1672 and 1673 triple output DC power supplies provide two variable outputs (0 – 32V/ 0 – 3A [mode

Page 22

6 Retain the original packing in case subsequent repackaging for return is required. Use of the original packing is essential. After the mechanical

Page 23

7 4 Controls and Indicators 4.1 Front Panel Figure 2 – Front Panel 1) Power switch 2) Negative output terminal of the fixed 5V/3A output 3)